IP Rating
IP20 (as defined in IEC 60529).
Ventilation Requirement
This oscilloscope uses fan to force cooling. Please make sure that the air intake and
exhaust areas are free from obstructions and have free air. When using the oscilloscope in
a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind
the instrument for adequate ventilation.
WARNING
Inadequate ventilation may cause temperature increase which would damage
the instrument. So please keep the instrument well ventilated during operation
and inspect the intake and fan regularly.
AC Power and Ground Connections
The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC power at 50/60
Hz (+/-5%), or single-phase 100 - 120 Vrms (+/-10%) AC power at 400Hz (+/-5%).
No manual voltage selection is required because the instrument automatically adapts to
line voltage.
Depending on the type and number of options and accessories (probes, PC port plug-in,
etc.), the SDS1xx2X-E can consume up to 25 W of power, the SDS1xx4X-E can consume up
to 25 W of power, the SDS1xx4X-U can consume up to 50 W of power.
Note: The instrument automatically adapts to the AC line input within the following
ranges:
Voltage Range:
90 - 264 Vrms
90 - 132 Vrms
Frequency Range:
47 - 63 Hz
380 - 420 Hz
The instrument includes a grounded cord set containing a molded three-terminal