SDS1000X-E&SDS1000X-U User Manual
241
WWW.SIGLENT.COM
“Single” and whether the trigger level exceeds the waveform range. If yes,
set the trigger level to the middle or set the mode to “Auto”.
Note: using AUTO could automatically finish the above setting.
6. The display of waveform is ladder-like:
1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
2) If the display Type is “Vectors”, the lines between the sample points may
cause ladder-like display. Set Type to “Dots” to solve the problem.
7. Fail to connect PC through USB:
Check the IO Setting in Utility to make sure whether the setting in USB Device
matches the device currently connected. If needed, restart the oscilloscope.
8. The USB storage device cannot be recognized:
1) Check whether the USB storage device can work normally.
2) Make sure whether the USB interface can work normally.
3) Make sure that the USB storage device being used is flash storage type.
This oscilloscope does not support hardware storage type.
4) Restart the instrument and then insert the USB storage device to check it.
5) If the USB storage device still cannot be used normally, please contact
SIGLENT.