SDS1000X-E&SDS1000X-U_User Manual - Page 141

For SDS1000X-E.

Loading ...
Loading ...
Loading ...
SDS1000X-E&SDS1000X-U User Manual
108
WWW.SIGLENT.COM
UART Trigger
This part shows a brief introduction and description for the operation of the UART
trigger.
1. Press the Setup button to enter the TRIGGER function menu.
2. Press the Type and select Serial.
3. Press the Protocol and select UART.
4. Press the Trigger Setting softkey to enter UART TRIG SET menu.
Figure 56 UART TRIG SET Menu
5. Press the Source Type softkey to select the source of trigger (RX or TX).
6. Press the Condition softkey and set up the desired trigger condition:
Start the oscilloscope will be triggered at the position of start bit.
Stop the oscilloscope will be triggered at the position of stop bits.
Data the oscilloscope will be triggered when found a byte which is
equal to (greater or less than) the specified data.
a. Press the Compare Type softkey and choose an equality qualifier (>, <
or =).
b. Press the Value softkey to set data’s value. Datas value is in range of
0x00 to 0xff.
ERROR if the parity check has been set, and the bit of parity check is
error, the oscilloscope will be triggered.
Figure 57 UART Trigger