SDS1000X-E&SDS1000X-U_User Manual - Page 122

For SDS1000X-E.

Loading ...
Loading ...
Loading ...
SDS1000X-E&SDS1000X-U User Manual
89
WWW.SIGLENT.COM
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified pattern.
The pattern trigger can be expanded to incorporate delays similar to other triggers.
Pattern durations are evaluated using a timer. The timer starts on the last edge
that makes the pattern “true”. Potential triggers occur on the first edge that
makes the pattern false, provided that the time qualifier criterion has been met.
The oscilloscope provides 4 patterns: logical AND, OR, NAND and NOR combination
of the channels. Each channel can set to low, high or invalid.
Do the following steps to set pattern trigger:
1. Press the
Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; turn the Universal Knob to select Pattern, and then push
down the knob to confirm.
3. Press each channel softkey to select Invalid, High or Low.
Low sets the pattern to low on the selected channel. A low is a voltage
level that is less than the channel's trigger level or threshold level.
High sets the pattern to high on the selected channel. A high is a voltage
level that is greater than the channel's trigger level or threshold level.
Don’t care sets the pattern to don't care on the selected channel. Any
channel set to don't care is ignored and is not used as part of the pattern.
However, if all channels in the pattern are set to Don’t care, the
oscilloscope will not trigger.
Adjust the trigger level for the selected analog channel by turning the
Trigger Level knob. Don’t care doesn’t need to set trigger level.
4. Press the Next Page softkey to enter the second page of the pattern trigger
menu.
Loading ...
Loading ...
Loading ...