Loading ...
Loading ...
Loading ...
70
Guidance and manufacturer’s declaration –
electromagnetic compatibility
The 7.0 R is intended for use in the electromagnetic environment specied below. The customer or the
user of the 7.0 R should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
guidance
RF emissions
CISPR 11
Group 1
The 7.0 R uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The 7.0 R is suitable for use in all
establishments, including domestic
establishments
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic elds
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey, a) should
be less than the compliance level in
each frequency range. B) Interference
may occur in the vicinity of equipment
marked with the following symbol:
Recommended separation distances between
portable and mobile RF communications equipment and the 7.0 R
The 7.0 R is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the 7.0 R can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the 7.0 R as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =
1,2
80 MHz to 800
MHz
d =
1,2
800 MHz to 2,5 GHz
d =
2,3
0,01
0,12
0,12
0,23
Dimensions
Length : 57” (145cm)
Width : 30” (77cm)
Height : 51” (130cm)
Weight
180.1 lbs. (81.7 kg)
Resistance
Constant and Isokinetic with
50 levels of eort.
Work load
5 watts up to 750 watts.
Disposal
Reference should be made to local regulations concerning the disposal of
this product at the end of useful life.
Certications
TUV listed to ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1, CAN/CSA-C22.2
No. 60601-1:14,
CE conformity to EN 60601-1 EMC, Compliance to EN 60601-1-2
Classication
Class I measuring, Type B, ordinary equipment, continuous operation. This
product is classed as ordinary equipment according to IEC/EN/UL60601-1