If a signal is lost for two consecutive decisions (roughly 30 seconds), the display will beep four times,
show HR SIGNAL LOST, show dashes in the BPM window and show the heart solid white. During this
HR signal lost period, the load will remain steady.
If after two further decisions (roughly an additional 30 seconds) the heartrate signal has not been
reestablished, continue to show HR SIGNAL LOST on the Text area. The program will reduce the
workload by 10 steps per minute until minimum workload or Step 1. After one minute at minimum
workload, go to Exiting Active Mode.
Cool Down
By default, Cool Down occurs at session time minus two minutes. Cool Down goes to 75% of existing
step, then drops by 1 step every 15 seconds for the remainder of Cool Down.