SHS800X&SHS1000X User Manual
8
WWW.SIGLENT.COM
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Set the switch to 10X on the probe.
2. Perform steps 1, 2, 3 and 4 of “
Function Inspection” in the previous section.
3. Check the waveforms displayed and compare them with the following:
Under Perfectly Over
Compensated Compensated Compensated
4. Use a nonmetallic driver to adjust the low-frequency compensation
adjustment hole on the probe until the waveform displayed is as the
“Perfectly compensated” in the figure above.
Multimeter Meter Pen
To avoid obtaining no measurements or unnecessary damage to the
SHS800X/SHS1000X, you should use the right jack when measuring current,
voltage and other measurement.