Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another as
shown in the picture below.
A positive runt pulse across through a lower threshold but not an upper
threshold.
A negative runt pulse across through an upper threshold but not a lower
threshold.
To trigger on runt pulse:
1. Press the Trigger Setup button on the front panel to enter the TRIGGER system
function menu.
2. Press the Type softkey; then turn the Universal Knob to select Runt and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1 or CH2 as the
trigger source.
4. Press the Polarity softkey to select Positive or Negative pulse to trigger.
5. Press the Limit Range softkey; turn the Universal Knob to select the desired
condition (<=, >=, [--,--] or --][--).
6. Press the Time Setting softkey, and then turn the Universal Knob to select the
desired value.
7. Press the Next Page softkey to enter the second page of the TRIGGER system
function menu. Press the Lower Upper softkey to select Lower or Upper
trigger level, and press the Trigger Level button and turn the Universal Knob
to set the position.