SHS800X&SHS1000X User Manual
176
WWW.SIGLENT.COM
Reference Position
The reference position setting determines the physical point that the oscilloscope
uses during vertical and horizontal scale changes. In some situations, it is more
convenient to use a fixed position on the display.
Press the Utility button on the front panel and then press the Reference Pos.
softkey to enter the Reference POS menu.
Press the Vertical softkey to select Fixed Offset or Fixed Position.
Fixed Position: The oscilloscope will keep the vertical offset level indicator
stationary when the vertical gain is changed.
Fixed offset: The oscilloscope will have the vertical offset level indicator move
with the actual voltage level when the vertical gain is changed.
Press the Horizontal softkey to select Fixed Delay or Fixed Position.
Fixed Position: when the timebase changes, the oscilloscope will keep the
horizontal offset indicator stationary.
Fixed Delay: when the timebase changes, the oscilloscope will have the
horizontal offset indicator move with the trigger point.