UTD1000L User Manual
type has been set up can the waveform displa
y
steady
.
3. Try to change Coupling to HF R eject to filter out
the high-frequency inteference noise in the
signal
.
1 Check if Trigger Mode in trigge r menu is
Normal or Single and trigger level has gone
beyond waveform ranges. If they are, set trigger
level to the middle or set trigger Mode to
A
UTO
2. Or you can press AUTO to automatically finis
h
the setup mentioned above
1. If AVG number go above 32times, it is normal for
the oscilloscopes to display at a relatively low
rate
.
2. Reduce AVG number
1. It is normal situation, it may b e caused by lower
horizont al ti
mebase, increasing the timebase
can improve the horizontal resolution and the
displa
y
2. If display type is VECTOR ,links between sample
points could probably result in stair-steppin
g
shape. Change display type to D ots can solve
the problem.
There is no display after pres sin
g
down RUN STOP butto
n
The display slows down after havin
g
applied AVG mode for a while:
The waveform display has a stair
-
stepping shape:
