Rohde & Schwarz NGP824 power supply

NGP800 series manual - Page 50

For NGP824.

PDF File Manual, 190 pages, Read Online | Download pdf file

NGP824 photo
Loading ...
Loading ...
Loading ...
Instrument Functions
R&S
®
NGP800
48User Manual 5601.5610.02 ─ 04
5. Press the required channel key ([Ch 1], [Ch 2], [Ch 3] or [Ch 4]) on the front panel.
The selected channel key is illuminated. See Figure 6-1.
6. Press the [Output] key on the front panel.
The R&S NGP800 outputs the set voltage of the selected channel and displays the
corresponding values in the home window.
For more information on the operation modes, see
Chapter 5.5, "Operation
Modes", on page 45.
Figure 6-2: Voltage and current settings in the instrument
6.2 Activating the Channels Output
The outputs of all the channels (Ch 1, Ch 2, Ch 3, Ch 4) can be switched on or off by
toggling the [Output] key on the front panel.
By default, the output is turned off when the instrument is switched on.
1. Press the required channel key.
Selected channel key (Ch 1, Ch 2, Ch 3, Ch 4) illuminates.
2. Press [Output] key.
The R&S NGP800 outputs the set voltage of the selected channel.
Depending on the operating mode, the font text in the channel display area shows
green in CV mode and red in CC mode.
See
Chapter 5.5, "Operation Modes", on page 45.
See also Chapter 5.4, "Output Power Auto Ranging", on page 44.
Multiple outputs can be turned on or off at the same time.
Activating the Channels Output
Loading ...
Loading ...
Loading ...