Remote Control Commands
R&S
®
NGP800
147User Manual 5601.5610.02 ─ 04
ARBitrary:BLOCk:ENDPoint?
Queries the number of data points of the block of arbitrary data.
Example:
INST OUT1
ARB:BLOC 1
ARB:BLOC:ENDP?
Return the number of data points for block 1 of Ch 1.
Usage: Query only
ARBitrary:BLOCk:FNAMe <>[, <>]
ARBitrary:BLOCk:FNAMe? <>[, <>]
Sets or queries the filename for block of arbitrary data.
Parameters for setting and query:
<filename> Filename of the QuickArb function.
<location> INT | EXT | DEF
INT
Internal memory
EXT
USB stick
DEF
Internal memory
Example:
INST OUT1
ARB:BLOC 1
ARB:BLOCK:FNAM "01.CSV"
ARB:BLOCK:FNAM? INT -> "01.CSV"
ARBitrary:BLOCk:REPetitions <>
Sets or queries the number of repetitions of the block of arbitrary data.
Parameters:
<repetitions> Repetition of the block of arbitrary data.
Example:
INST OUT1
ARB:BLOC 1
ARB:BLOC:REP 0
Set repetition of infinity to block 1 of Ch 1.
ARBitrary[:STATe] <>[, <Channel list>]
ARBitrary[:STATe]? [<Channel list>]
Sets or queries the QuickArb function for the previous selected channel.
Parameters:
<state> 1
QuickArb function is activated.