Rohde & Schwarz NGP824 power supply

NGP800 series manual - Page 131

For NGP824.

PDF File Manual, 190 pages, Read Online | Download pdf file

NGP824 photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
NGP800
129User Manual 5601.5610.02 ─ 04
Example:
FUSE:TRIP:CLE? (@1)
Queries OCP reset state at channel 1.
Example: See
Example "Configuring fuses" on page 127.
Usage: Setting only
FUSE:DELay:INITial <New value for voltage>[, <Channel list>]
FUSE:DELay:INITial? [<Channel list>]
Sets the initial fuse delay time once output turns on.
Parameters:
<duration> <numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value for initial fuse delay.
MIN | MINimum
Min value for initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range: 0.00 to 60.00
*RST: 0
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Example:
FUSE:DEL:INIT? (@1)
Queries initial fuse delay time at channel 1.
Example: See
Example "Configuring fuses" on page 127.
FUSE:DELay[:BLOWing] <New value for voltage>[, <Channel list>]
FUSE:DELay[:BLOWing]? [<Channel list>]
Sets the fuse delay time.
Parameters:
<duration> <numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value for the initial fuse delay.
MIN | MINimum
Min value for initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range: 0.00 to 10.00
*RST: 0
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Configuration Commands