Remote Control Commands
R&S
®
NGP800
134User Manual 5601.5610.02 ─ 04
<numeric value>
Numeric value for the overvoltage protection value in V.
MIN | MINimum
Minimum value for the overvoltage protection value at 0.000 V.
MAX | MAXimum
Maximum value for the overvoltage protection value at 64.050 V.
Range: 0.000 to 64.050
*RST: 64.050
Default unit: V
Parameters for setting and query:
<Channel list> <list>
Example:
VOLT:PROT:LEV? (@1)
Queries overvoltage protection value at channel 1.
Example: See
Example "Configuring the overvoltage protection"
on page 132.
[SOURce:]VOLTage:PROTection:TRIPped? [<Channel list>]
[SOURce:]VOLTage:PROTection:TRIPped?? [<Channel list>]
Queries the OVP state of the selected channel.
Parameters:
<Channel list> <list>
Example:
VOLT:PROT:TRIP?
Response 1, the OVP is tripped.
Response 0, the OVP is not tripped.
Example:
VOLT:PROT:TRIP? (@1)
Queries OVP state at channel 1.
Usage: Query only
See
Example "Configuring the overvoltage protection" on page 132.
7.5.10 OPP Setting
The POWer:PROTection subsystem contains the commands for setting the over-
power protection parameters for the output channels. The default unit is W.
Configuration Commands