Chapter 2 Front Panel Operation RIGOL
User’s Guide for DSA700 Series
2-33
Pass/Fail
DSA700 supports the Pass/Fail test function. In this function, the measured curve
is compared with the pre-edited curve. If the related rules are met, the result is
"Pass"; or else is "Fail". The measurement interface is as shown in the figure
below.
Figure 2-11 Interface of Pass/Fail Test
Result:
The results of Pass/Fail test contain the pass number and the pass ratio.
1. Switch
Enable or disable the Pass/Fail test function and the default is "Off". When the
function is enabled, the system changes to split-screen display. At this point,
T2 and T3 are used for marking the upper and lower limits separately.
2. Settings
Limit: select the desired limit line (upper or lower) for editing. Then,
press Edit to set the limit conditions.
Test: enable or disable the test function. Each limit line has its own