RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA700 Series
2-26
side of the screen.
Single Continuous
1. Single
Set the sweep mode to "Single". The number 10 on the parameter icon
denotes the current sweep number.
2. Cont
Set the sweep mode to "Cont". The character Cont on the parameter icon
denotes the analyzer is sweeping continuously.
Key Points:
If the instrument is in single sweep mode and no measurement function is
enabled, press this key and the system will enter continuous sweep mode and
sweep continuously if the trigger conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on,
press this key and the system will enter continuous sweep mode and measure
continuously if the trigger conditions are satisfied.
In continuous sweep mode, the system will send a trigger initialization signal
automatically and enter the trigger condition judgment directly after each
sweep.
Enter continuous sweep mode and
wait for trigger condition
Sweep again and
wait for trigger
condition
Select Cont
Single sweep status
Y
N
Figure 2-9 Process of Continuous Sweep