Rigol DSA-705 spectrum analyser

Rigol DSA 700 manual - Page 204

For DSA-705.

PDF File Manual, 209 pages, Read Online | Download pdf file

DSA-705 photo
Loading ...
Loading ...
Loading ...
RIGOL Chapter 5 Specifications
User’s Guide for DSA700 Series
5-10
Electromagnetic Compatibility and Safety
EMC
in line with EMC instruction (2014/30/EU),
in line with or exceed IEC61326-1: 2013/EN61326-1: 2013 Group 1 Class A standard
CISPR 11/EN 55011
IEC 61000-4-2:2008/EN
61000-4-2
±4.0 kV (contact discharge), ±8.0 kV (air discharge)
IEC 61000-4-3:2002/EN
61000-4-3
3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz to 2 GHz); 1 V/m
(2.0 GHz to 2.7 GHz)
IEC 61000-4-4:2004/EN
61000-4-4
1 kV power lines
IEC 61000-4-5:2001/EN
61000-4-5
0.5 kV (phase to neutral); 0.5 kV (phase to PE); 1 kV
(neutral to PE)
IEC 61000-4-6:2003/EN
61000-4-6
3 V, 0.15-80MHz
IEC
61000-4-11:2004/EN
61000-4-11
voltage dip: 0% UT during half cycle; 0% UT during 1 cycle;
70% UT during 25 cycles
short interruption: 0% UT during 250 cycles
Electrical safety
IEC 61010-1:2010 (Third Edition)/EN 61010-1:2010,
UL 61010-1:2012 R4.16 and CAN/CSA-C22.2 NO.
61010-1-12+ GI1+ GI2
Dimensions
(W × H × D)
361.6 mm × 178.8 mm × 128 mm
(14.2 in × 7.0 in × 5.0 in)
Weight
DSA705 DSA710
Standard 4.25 kg (9.4 lb)
Calibration Interval
Recommended calibration
interval
1 year
Loading ...
Loading ...
Loading ...