Specifications
DSO5000 Series Digital Storage Oscilloscope User Manual 64
CH1
CH2
1div from DC to
10MHz;
1.5div from 10MHz to
Full
1.5div from 10MHz
to 100MHz;
2div from 100MHz to
Full
EXT
200mV from DC to
100MHz
200mV from DC to
100MHz; 350mV
from 100MHz to
200MHz
EXT/5
1V from DC to
100MHz
1V from DC to
100MHz;
1.75V from 100MHz
to 200MHz
AC Attenuates signals below 10Hz
HF Reject Attenuates signals above 80kHz
LF Reject
Same as the DC-coupled limits for frequencies above
150kHz; attenuates signals below 150kHz
Source Range
CH1, CH2 ±8 divisions from center of screen
EXT ±1.2V
Trigger Level Range
EXT/5 ±6V
Source Accuracy
CH1CH2
0.2div × volts/div within ±4 divisions from center of
screen
EXT ± (6% of setting + 40mV)
Trigger Level
Accuracy, typical
(Accuracy is for
signals having rise
and fall times 20ns)
EXT/5 ± (6% of setting + 200mV)
Set Level to 50%,
typical
Operates with input signals 50Hz
Note: Bandwidth reduced to 6MHz when using a 1X probe.
Video Trigger Type Source Range
CH1, CH2
Peak-to-peak amplitude of 2
divisions
EXT 400mV
EXT/5 2V
Signal Formats and
Field Rates, Video
Trigger Type
Supports NTSC, PAL and SECAM
broadcast systems for any field or
any line
Holdoff Range 100ns to 10s
Pulse Width Trigger
Pulse Width Trigger
Mode
Trigger when < (Less than), > (Greater than), = (Equal), or (Not Equal);
Positive pulse or Negative pulse