Basic Operation
DSO5000 Series Digital Storage Oscilloscope User Manual 40
Functions Settings
Acquire Mode Adjusted to Normal or Peak Detect
Cursor Off
Display Format Set to YT
Display Type Set to Vectors for an FFT spectrum; otherwise, unchanged
Horizontal Position Adjusted
SEC/DIV Adjusted
Trigger Coupling Adjusted to DC, Noise Reject or HF Reject
Trigger Holdoff Minimum
Trigger Level Set to 50%
Trigger Mode Auto
Trigger Source Adjusted; Autoset can not be used for the EXT TRIG signal
Trigger Slope Adjusted
Trigger Type Edge
Trigger Video Sync Adjusted
Trigger Video Standard Adjusted
Vertical Bandwidth Full
Vertical Coupling
DC (if GND was chosen before); AC for the video signal; otherwise,
unchanged
VOLTS/DIV Adjusted
The Autoset function examines all channels for signals and displays corresponding waveforms.
Autoset determines the trigger source according to the following conditions.
If multiply channels get signals, the oscilloscope will use the channel with the lowest
frequency signal as the trigger source.
If no signals are found, the oscilloscope will use the lowest-numbered channel displayed
in Autoset as the trigger source.
If no signals are found and no channels are displayed, the oscilloscope will display and
use Channel 1 as the trigger source.
Sine Wave:
When you use the Autoset function and the oscilloscope determines that the signal is similar to a
sine wave, the oscilloscope displays the following options.
Sine Wave Options Details
Multi-cycle Sine
Display multiple cycles that have appropriate vertical and
horizontal scales.
Single-cycle Sine Set the horizontal scale to display about one cycle of the waveform.
FFT
Convert the input time-domain signal to its frequency components
and display the result as a graph of frequency versus amplitude
(spectrum). Since it is a math calculation, see Section 5.3.1 Math
FFT for more information.
Undo Setup Let the oscilloscope recall the previous setup.
Square Wave or Pulse:
When you use the Autoset function and the oscilloscope determines that the signal is similar to a
square wave or pulse, the oscilloscope displays the following options.
Square Wave Options Details
Multi-cycle Square
Display multiple cycles that have appropriate vertical and
horizontal scales.
Single-cycle Square Set the horizontal scale to display about one cycle of the waveform.