Specifications
DSO5000 Series Digital Storage Oscilloscope User Manual 63
Vertical
A/D Converter
8-bit resolution,
each channel sampled simultaneously
VOLTS/DIV Range 2mV/div to 5V/div at input BNC
Position Range
2mV/div to 200mV/div, ±2V
>200mV/div to 5V/div, ±50V
Analog Bandwidth in
Normal and Average
modes at BNC or with probe, DC
Coupled
2mV/div to 20mV/div, ±400mV
50mV/div to 200mV/div, ±2V
500mV/div to 2V/div, ±40V
5V/div, ±50V
Selectable Analog Bandwidth Limit,
typical
20MHz
Low Frequency Response (-3db) 10Hz at BNC
DSO5062C DSO5102C DSO5102B DSO5202B
Rise Time at BNC, typical
<5.8ns
3.5ns 3.5ns
<1.8ns
DC Gain Accuracy
±3% for Normal or Average acquisition mode, 5V/div to
10mV/div
±4% for Normal or Average acquisition mode, 5mV/div
to 2mV/div
Measurement Type: Average of 16 waveforms with
vertical position at zero
Accuracy: ± (3% × reading + 0.1div + 1mV) when
10mV/div or greater is selected
DC Measurement Accuracy,
Average Acquisition Mode
Measurement Type: Average of 16 waveforms with
vertical position not at zero
Accuracy: ± [3% × (reading + vertical position) + 1% of
vertical position + 0.2div]
Add 2mV for settings from 2mV/div to 200mV/div; add
50mV for settings from 200mV/div to 5V/div
Volts Measurement Repeatability,
Average Acquisition Mode
Delta volts between any two averages of 16 waveforms
acquired under same setup and ambient conditions
Note: Bandwidth reduced to 6MHz when using a 1X probe.
Trigger
Coupling Sensitivity Trigger Sensitivity
(Edge Trigger Type)
DC
Source
DSO5062C
DSO5102C
DSO5102B
DSO5202B
Loading ...