Basic Operation
DSO1000S Series HandHeld Oscilloscope User Manual 37
arrange the sample points by the time they appear, then recover waveforms.
5.5.4 UTILITY
Push the UTILITY key to display the Utility Menu as follows.
Options
Comments
System Info
Display the software and hardware versions, serial number and some other
information about the oscilloscope.
Update
Insert a USB disk with upgrade program and the disk icon at the top left corner is
highlighted. Press the Update Program button and the Software Upgrade dialog
pops up.
Self Cal
Press this option and the Self Calibration dialog pops up. Press F6 to perform the
self calibration.
System
Enter the system setup page.
ShutDown
Ener the Shutdown page.
Video
Enter the playing video page.
Probe CK
Enter the Probe-CK page.
Pass/Fail
Enter the Pass/Fail page.
Record
Enter the Record page.
Filter
Enter the Filter page.
Display
Enter the Display page.
Acquire
Enter the Acquire page.
DMM
ON/OFF
Frequency
ON/OFF
More
To check the system information.
Self Cal: The self calibration routine can optimize the precision of the oscilloscope to fit the
ambient temperature. To maximize the precision, you should perform the self calibration once the
ambient temperature changes by 5 or more. Follow the instructions on the screen.
Tip: Press any menu button on the front panel to remove the status display and enter a
corresponding menu.
5.5.5 CURSOR
The Cursor Menu.
Options
Settings
Comments
Type
Off
Voltage
Time
Select a measurement cursor and display it.
Voltage measures amplitude while Time measures time and
frequency.
Source
CH1
CH2
MATH
REFA
Select a waveform to take the cursor measurement.
Use the readouts to show the measurement.