MKF / MKFT (E5) 06/2020 page 78/176
10.6.4 Setpoint entry
Select the field Temperatureand enter the desired temperature setpoint.
Setting range: MKF: -50 °C up to 180 °C, MKFT: -80 °C up to 180 °C.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field Humidityand enter the desired humidity setpoint.
Setting range: 0% r.h. up to 100% r.h.
Confirm entry with Confirm icon. The controller returns to the section view.
MKF 56: Select the field Fanand enter the desired fan speed setpoint.
Setting range: 30% up to 100% fan speed.
Confirm entry with Confirm icon. The controller returns to the section view.
10.6.5 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line Humidity offserves to turn off the humidity.
Operation line Idle modeactivates / deactivates the operating mode Idle mode”.
Operation lines Switching output 1” up to “4” can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (7) and (8) (chap. 20).
Operation line Compr. air dryerserves to activate the compressed air dryer (option, chap. 21.6).
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air con-
nection (for options compressed air connection, chap. 21.5, or compressed air dryer, chap. 21.6)
Operation line Anti-condensationserves to switch the anti-condensation protection (chap. 19).
The other operation lines are without function.
Select the desired program and section. You can set the operation lines in the “Functions on/off” field.
For details please refer to chap. 9.7.3.