MKF / MKFT (E5) 06/2020 page 29/176
Do not install or operate the chamber in potentially explosive areas.
DANGER
Danger of explosion due to combustible dusts or explosive mixtures in the vicinity
of the chamber.
Serious injury or death from burns and / or explosion pressure.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Ambient conditions
Permissible ambient temperature range during operation: +18 °C / 64.4 °F to +32 °C / 89.6 °F. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of +22 °C +/- 3 °C to which the specified technical data relate. For other am-
bient conditions, deviations from the indicated data are possible.
Permissible ambient humidity: 70 % r.h. max., non-condensing.
When operating the chamber at temperature set-points below ambient temperature, high ambient hu-
midity may lead to condensation on the chamber.
Installation height: max. 2000 m / 6.6 ft. above sea level.
Minimum distances:
Distance between each chamber when placing several chambers of the same size side by side: 250
mm / 9.84 in
Lateral wall distance: 200 mm / 7.87 in
Wall distance rear: 300 mm / 11.81 in
Chambers with optional water cooling (without options compressed air dryer and / or voltage and fre-
quency changer): Wall distance rear 100 mm / 3.94 in.
Chambers with optional compressed air dryer: Wall distance rear approx. 1 m / 3.28 ft so that it is pos-
sible to read the status display of the compressed air dryer on the chamber rear.
Chambers with voltage and frequency changer: rear wall distance of the alternating climate chamber
approx. 1 m / 3.28 ft to set up the voltage and frequency changer
To completely separate the chamber from the power supply, you must disconnect the power plug. In-
stall the chamber in a way that the power plug is easily accessible and can be easily pulled in case of
danger.
Spacing above the chamber: 100 mm / 3.94 in
NOTICE
Danger by stacking.
Damage to the chambers.
Do NOT place the chambers on top of each other.
Other requirements
A water tap (1 bar to 10 bar) is necessary for the installation of the humidification system. If no suitable
house water connection is available, you can manually supply water by filling the water can (MKF 56:
optional external water can, chap. 4.2.2, 21.9, MKF/MKFT 115, 240, 720: internal water can chap. 4.2.3).
To avoid any possible water damage, provide a floor drain at the location of the device. Select
a suitable installation site to avoid any consequential damage by splashing water.