Binder MKF 240 Dynamic climate chamber with humidity control, -liter capacity

User Manual - Page 56

For MKF 240.

PDF File Manual, 177 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MKF / MKFT (E5) 06/2020 page 56/176
In the field Program typeselect Timer program”.
Select the field Program durationand enter the desired program duration. Press the Confirm icon.
Select the field Program start and enter the desired start time of the program in the Program start
entry menu. Press the Confirm icon. The program delay time until program start begins.
Normal display.
Information on the bottom of the screen indicates the
currently running program and the time already passed.
The grey bar shows how much time of the whole time is
elapsed.
8.1.1 Performance during program delay time
During the configured program delay time until program start, the controller equilibrates to the current
setpoints of Fixed value operation mode. Modifications of these setpoints are possible but become effec-
tive only after the timer program is finished. When the configured moment for program start is reached,
the program delay time ends and the program starts running. The controller equilibrates to the values
which had been active during program start.
8.2 Stopping a running timer program
8.2.1 Pausing a running timer program
Press the Program pause icon to interrupt the program.
The program is paused. The program runtime stops running down, the time display flashes.
There are the following options:
Press the Program start icon to continue the program
Press the Cancelling icon to cancel the program
8.2.2 Cancelling a running timer program
Press the Program cancelling icon to cancel the program.
A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be
cancelled.
After confirming the message the controller changes to Fixed value operation mode. Temperature and
humidity will then equilibrate to the setpoints of Fixed value operation mode.
8.3 Performance after the end of the program
After the end of the program the message “Device
changes to fixed value operation mode” appears on the
screen.
Press the Confirm icon.
After confirming the message the controller changes to Fixed value operation mode. Temperature and
humidity will then equilibrate to the setpoints of Fixed value operation mode.