Water connections
With the optional water cooling the chamber is supplied with cooling water via a freshwater pipe (max.
inlet temperature: 10 °C / 50 °F).
Connection of cooling water inlet: please refer to chap. 4.4.
Connection of cooling water outlet: please refer to chap. 4.3.
21.8 Object temperature display with flexible Pt 100 temperature sensor (option)
The object temperature display enables the determination of the actual temperature of the charging mate-
rial during the whole process. The object temperature is measured via a flexible Pt100 temperature sen-
sor and can be viewed on the controller display. You can immerse the sensor top protective tube of the
flexible Pt 100 into liquid substances.
Normal display with object temperature display
(sample values)
The object temperature data are put out together with the data of the temperature controller to the inter-
face as second measuring channel and can be documented by the APT-COM™ 4 Multi Management
Software (option, chap. 21.1) developed by BINDER.
Technical data of the Pt 100 sensor:
Three-wire technique
Class B (DIN EN 60751)
Temperature range up to 320 °C / 608°F
Stainless steel protective tube with a length of 45 mm / 1.78 in, material no. 1.4501