MKF / MKFT (E5) 06/2020 page 128/176
Cable connection between the pump and the climatic chamber
Connect the cable plug from the cable connection (21) on the pump to the
socket (18) at the rear of the chamber.
The socket (18) is marked with a sticker:
Socket (18)
Hose connections between the pump and the climatic chamber
Plug the freshwater hose into the hose connection (22) on the pump and secure it with a hose clamp. You
can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwater con-
nection “IN” (15) at the rear of the chamber.
When the freshwater can is empty, the message Freshwater supplywill be displayed on the controller
(chap. 11.1.3), the buzzer sounds, and the humidification module turns off. After acknowledging the
alarm, the humidification module tries to fill up and start operating.
To guarantee humidification during 24 hours even at high humidity set-points with manual
water supply, we recommend filling the freshwater can (option) at the end of each day.
21.9.2 Connecting the wastewater can
Hose connection
Plug the wastewater hose to the hose connection (23) of the wastewater can and secure it with a hose
clamp. You can use a part of the standard supplied water hose.
Hose connection (23)
of the wastewater can
Figure 26: Wastewater can
Plug the free hose edge to the wastewater connection “OUT” (14) at the rear of the chamber and secure it
with a hose clamp.
For emptying the wastewater can disconnect the hose first.