Loading ...
Loading ...
Loading ...
KBF / KBF-UL + KMF (E6) 04/2020 page 32/163
Figure 11: Assembly of the safety kit (hose burst protection and reflux protection devices)
Release of the reflux protection device:
In case the hose burst protection device interrupts the water supply, first find the reason and remove it as
necessary. Close the water tap. Release the valve by a half left-turn of the upper knurled part. You can
hear the release of the valve as a clicking noise. Tighten the burst protection device against the water tap
by a right turn. Open the water tap slowly afterwards.
Maintenance of the assembly of hose burst protection and reflux protection devices:
Calcification can impair the function of both valves. We recommend an annual inspection by a skilled
plumber. The plumber should remove the safety kit with the reflux protection device to check both valves
by hand for proper function and calcification or blockage.
NOTICE
Danger of impairment of the valve function by calcification.
Damage to the chamber.
Have a plumber inspect the valve annually.
Remove calcifications by citric acid or acetic acid solutions.
Continue by testing the function and tightness of the mounted chamber
Check: Quickly open the water tap while there is no chamber connected the valve should cut off the
water flux without any delay.
Loading ...
Loading ...
Loading ...