KBF / KBF-UL + KMF (E6) 04/2020 page 122/163
(2) Cable connections
Connect the plug of the cable to the socket (10) at the rear of the chamber.
The socket (10) is marked with a sticker:
Socket (10)
(14)
(13)
Figure 25: Connections at the chamber rear
(3) Hose connections
Plug the freshwater hose into the hose connection (16) above the freshwater can and secure it with
a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwater
connection “IN” (13) at the rear of the chamber.
When the freshwater can is empty, the message Freshwater supplywill be displayed on the controller
(chap. 11.1.3), the buzzer sounds, and the humidification module turns off. After acknowledging the
alarm, the humidification module tries to fill up and start operating.
To guarantee humidification during 24 hours even at high humidity set-points with manual
water supply, we recommend filling the freshwater can (option) at the end of each day.