KBF / KBF-UL + KMF (E6) 04/2020 page 25/163
3.2 Guidelines for safe lifting and transportation
The front castors of the chambers size 240, 720 and 1020 can be blocked by brakes. After operation,
please observe the guidelines for temporarily decommissioning the chamber (chap. 22.2). Please move
the chambers with castors only when empty and on an even surface, otherwise the castors may be dam-
aged.
CAUTION
Risk of injury and damages by lifting heavy loads and by sliding or tilting of the
chamber due to improper transportation.
Injuries, damage to the chamber.
Transport the chamber in its original packaging only.
For moving or shipping, secure the constant climate chamber with transport straps.
Do NOT lift or transport the chamber using the door, the handle, or the lower housing.
Lift chambers size 115 at the four lower corners with the aid of 4 people
Lift chambers size 240 at the four lower corners with the aid of 6 people or with a fork
lifter. Set the fork lifter only from the front or rear in the middle of the chamber.
Lift the chambers sizes 720 and 1020 using technical devices (fork lifter). Set the fork
lifter only from the front or rear in the middle of the chamber.
Do NOT set the fork lifter from the chamber side.
You can order transport packing for moving or shipping purposes from BINDER service.
Permissible ambient temperature range during transport:
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
After BINDER Service has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F.
With temperatures below +3 °C / 37.4 °F, water must be completely removed from the humidifying sys-
tem.
NOTICE
Danger of freezing in the steam generator when transporting the chamber below
+3 °C / 37.4 °F with filled steam humidifying system.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C / 37.4 °F.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 22.2).
Permissible ambient temperature range during storage:
If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.
After BINDER Service has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F