KBF / KBF-UL + KMF (E6) 04/2020 page 132/163
Fault description
Possible cause
Required measures
Controller (continued)
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do
NOT program tolerance limits in
order to permit maximum heating,
refrigerating, or humidification
speed.
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting “ramp”
is incomplete.
When programming with setting
“ramp”, define the end value of the
desired cycle by adding an addi-
tional section with a section time of
at least one second.
Ramp temperature transitions
are only realized as steps.
Setting “step” has been selected. Select setting “ramp”.
Humidity alarm when operating
without water connection.
Humidity control turned on.
Turn off humidity control (chap.
6.3).
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the
alarm state continues, contact
BINDER service.
Alarm message:
- - - - or <-<-< or >->->
Sensor rupture between sensor
and controller or Pt 100 sensor
defective.
Contact BINDER service.
Short-circuit.
Miscellaneous
Impaired valve function of hose
burst protection.
Calcification.
Remove calcifications by citric acid
or acetic acid solutions (chap.
4.3.4).
Have a plumber inspect the valve.
21.4 Sending the chamber back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number (RMA number) that has previously been issued to you. An
authorization number will be issued after receiving your complaint either in writing or by telephone prior
to your sending the BINDER product back to us. The authorization number will be issued following receipt
of the information below:
BINDER product type and serial number
Date of purchase
Name and address of the dealer from which you bought the BINDER product
Exact description of the defect or fault
Complete address, contact person and availability of that person
Exact location of the BINDER product in your facility
A contamination clearance certificate (chap. 26) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recognized
or be recorded clearly in the delivery documents.
For security reasons we cannot accept a chamber delivery if it does not carry an authorization
number.
Return address:
BINDER GmbH
Abteilung Service
Gänsäcker 16
78502 Tuttlingen
Germany