FLUKE 2680A-APSW/U Data Application Software Upgrade from V5 or below

User Manual for the Fluke 2680A-APSW/U Data Application Software Upgrade from V5 or below - Page 21

For 2680A-APSW/U.

PDF File Manual, 73 pages, Read Online | Download pdf file

2680A-APSW/U photo
Loading ...
Loading ...
Loading ...
2680A/2686A
Getting Started Guide
12
Reviewing and Setting the General Network Socket Port
Perform the procedure below to review or set the general network Socket Port
(1024 to 65535). The default is 04369. In order to communicate with each other, a
host computer and an device must use the same socket port number. (See Figure 4
for examples.)
General Network Socket Port Enter the Socket Port supplied by your
network administrator.
Table 5. Reviewing and Setting the General Network Socket Port
Press the COMM key to review the network settings, or press and hold the
COMM key for 3 seconds to set the network settings.
Press the up/down arrow keys until nEt (Network) appears in the primary
display (comm appears in the secondary display).
Press the ENTER key. nEt appears in the secondary display and ISo (isolated
network) or gEn (general network) is in the primary display.
To set the network type, press the up/down arrow keys to
select gEn.
Press the ENTER key. (Pressing any other function key will cancel set
operations.) This displays the current Socket Port.
To set the socket port, press the left/right arrow keys to select the desired
digit position (highlighted).
Press the up/down arrow keys to select the desired number, 0 to 9, for the
positioned Port digit. In this manner, select all Port digits.
Press the ENTER key. This displays the first digit of the Internet Protocol
address segment IP:0.
Press the ENTER key to enter the settings and exit the procedure. (Pressing
any other function key will cancel set operations.).
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...