FLUKE 1625-2 GEO Earth Ground Tester, -10 to 50°C

User Manual for the Fluke 1625-2 GEO Earth Ground Tester, -10 to 50°C - Page 47

For 1625-2.

PDF File Manual, 52 pages, Read Online | Download pdf file

1625-2 photo
Loading ...
Loading ...
Loading ...
1625-2
Users Manual
42
Specifications
Temperature Range
Operating: 0 °C to +35 °C (+32 °F to +95 °F)
Storage: -30 °C to +60 °C (-22 °F to +140 °F)
Temperature coefficient: ±0.1 % of rdg / °C (below 18 °C and above 28 °C)
Operating humidity: <95% RH non-condensing
Operating altitude: 2000 m
Climatic Class: C1 (IEC 654-1) -5 °C to +45 °C, 5% to 95% RH
Protection t
y
pe
Case: IP56
Battery door: IP40
Electromagnetic compatibility: Complies with IEC61326-1: Portable
Safety: Complies with IEC 61010-1: CAT None, Pollution Degree 2
Measurement time: 6 seconds, typical
Maximum overload: 250 V
rms
(in association with misuse)
Batteries: 6 x 1.5v, AA LR6 Alkaline
Battery life span: >3000 measurements, typical, RH + RE < 1 kOhm
>6000 measurements, typical, RH + RE > 10 kOhm
Dimensions: 240 mm x 180 mm x 110 mm (9.5 in x 7.1 in x 4.4 in)
Weight with batteries: 1.52 kg (3.35 lb)
Memory: Internal memory storage up to 1500 records
accessible via USB port
Interference Voltage DC + AC (U
ST
) Measurement
Measuring method: fullwave rectification
Measuring
Range
Display
Range
Resolution Frequency Range Accuracy
1…50 V 0.0…50 V 0.1 V DC/AC 45…400 Hz sine ± (5% of rdg +5 digit)
Measuring sequence: approximately 4 measurements/s
Internal resistance: approximately 1.5 MΩ
Maximum overload: U
rms
= 250 V
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...