Earth/Ground Tester
Display
11
Table 5. Display Descriptions (cont.)
Function Displays Condition Note
edw035.eps
Aux. current spike-
resistance is being
tested.
Wait for test result.
edw036.eps
Earth resistance is
being tested.
Wait for test result.
ST
edw037.eps
Measuring circuit of
earth and
auxiliary-earth
electrode
disconnected.
Check lead connection
on earth spikes
measuring lead might
be defective.
ST
edw038.eps
Measuring circuit of
earth and probe
electrode
disconnected.
Check lead connection
on earth spikes
measuring lead might
be defective.
edw039.eps
Maximum allowable
error exceeded
because of too high
sense or aux earth
spike resistance.
Try to moisten soil or
connect 2nd aux earth
spike in parallel.
After
"START"
edw040.eps
Measuring range
exceeded.
Measured value is
higher than 300 kΩ.
edw041.eps
Display of measured
value exceeds LIMIT.
Measured value is
higher than set LIMIT.