Earth/Ground Tester
Advanced Operation
37
How to Change All Data Settings with Personalized CODE
With this function (FM, UM-Limit, Limit, beeper, ratio, R*) limit and set values can be
programmed which keeps them in memory even if the instrument is switched ON/OFF.
This feature enables the operator to create an instrument setup with customer defined
settings according to the specific need.
Table 7 shows the settings that can be made only in the respective functions:
Table 7. Data Settings
Function Parameter Setting range Standard presetting
F
M
(AFC/94/105/111/128) Hz AFC
R
E
3pole U
M
48 V/20 V 48 V
R
K
0.000 Ω ... 29.99 Ω 0.000 Ω
and LIMIT On / Off Off
R
E
LIMIT 0.000 Ω ... 999 kΩ 999 kΩ
R
E
4pole (warning sound) On/Off Off
R* On/Off Off
F
M
(AFC/94/105/111/128) Hz AFC
R
E
3pole U
M
48 V/20 V 48 V
R
K
0.000 Ω ... 29.99 Ω 0.000 Ω
and I ( ratio ) 80 ... 1200 1000
R
E
4pole LIMIT On/Off Off
R
E
LIMIT 0.000 Ω ... 999 kΩ 999 kΩ
(warning sound) On/Off Off
R* On/Off Off
F
M
(AFC/94/105/111/128) Hz AFC
R
K
0.000 Ω ... 29.99 Ω 0.000 Ω
R~ LIMIT On/Off Off
R ~ LIMIT 0.000 Ω ... 999 kΩ 999 kΩ
(warning sound) On/Off Off
R
2pole R
K
0.000 Ω ... 29.99 Ω 0.000 Ω
and LIMIT On/Off Off
4pole R LIMIT 0.000 Ω ... 9,99 kΩ 9.99 kΩ
(Warning sound) On/Off Off