Chapter 10 Pass/Fail Test RIGOL
DS1000Z User Guide 10-3
Test and Output
You can set the output form of the test results by the following method.
Press Stat.Disp to select "ON" or "OFF". When "ON" is selected, the test results will
be displayed at the upper right corner of the screen as shown in the figure below.
Press Stat.Reset to clear the current data and perform statistics on the test results
again.
Press StopOnFail to select "ON" or "OFF".
ON: when failed waveforms are detected, the oscilloscope will stop the test and
enter the "STOP" state. At this point, if Stat.Disp is turned on, the last statistic
results will be displayed on the screen; if Aux Out is turned on, a pulse is output
from the [Trigger Out] connector on the rear panel.
OFF: the oscilloscope will continue with the test even though failed waveforms
are detected. The test results on the screen will update continuously. If Aux
Out is turned on, a pulse will be output from the [Trigger Out] connector on
the rear panel each time a failed waveform is detected.
Press Output to select " " or " ".
: when failed waveforms are detected, the instrument displays the statistic
results and outputs a pulse.
: when failed waveforms are detected, the instrument displays the statistic
results and outputs a pulse and the beeper sounds (not related to the on/off
state of the sound).
Press AuxOut to quickly turn "ON" or "OFF" the output of test results from the
[Trigger Out] connector on the rear panel. When a failed waveform is detected, the
oscilloscope outputs a negative pulse from the [Trigger Out] connector on the rear
panel; when no failed waveform is detected, the instrument will output low level
continuously.