3 0 2 int.siglent.com
29 Bode Plot
Overview
The SDS6000A supports an automatic Bode plot function. This function
provides a frequency response curve of the device-under-test as well as the
interface for output sweep parameter control and data display settings. At this
time, either the SIGLENT SAG1021I arbitrary waveform generator or one of a
SIGLENT SDG series arbitrary function generators are supported. During the
sweep, the oscilloscope configures the generator output frequency and
amplitude and then compares the input signal to the output of the DUT. Gain
(G) and phase (P) are measured at each frequency and plotted on the
frequency response Bode plot. When the loop response analysis is complete,
you can move the markers on the chart to see the gain and phase values
measured at each frequency point. You can also adjust the scale and offset
settings for the amplitude and phase plots.
Touch Analysis>Bode Plot to recall Bode plot dialog box: