Frame The oscilloscope triggers on the frame.
Set Frame header indicators: Payload preamble indicator, null frame
indicator, sync frame indicator, startup frame indicator.
Touch ID to set the frame ID by the universal knob or virtual keypad.
The range of ID is 0x000 to 0x7ff.
Touch Compare Type to select "=", ">" or "<". When setting the
Compare Type as "=", the repetition factor is also required.
Touch Cycle to set the cycle count by the universal knob or virtual
keypad. The range of data values is 0 to 63.
Symbol The oscilloscope triggers on CAS/MTS (Conflict Avoidance
Symbol/ Media access Test Symbol) or WUS (Wake-up Symbol)
Errors The oscilloscope will trigger when a data error happens, including
errors on FSS, BSS, FES, Header CRC, and Frame CRC.
17.7.3 FlexRay Serial Decode
The configuration of FlexRay decoding is similar to that of I2C decoding.
On the bus:
The signatures (CAS/MTS, WUP) are displayed in yellow-green.
TSS transmission start sequence, displayed in yellow-green. The null
frame indicator, the Sync frame indicator, and the Startup frame
indicator are displayed in the frame and displayed in pink.
The ID is displayed in the frame and is displayed in green.
PL (Valid Data Length) is displayed in frames, in words, in light
yellow.