SDS6000A Series Digital Oscilloscope User Manual
int.siglent.com 15
Note: Degree of Pollution II refers to a working environment that is dry and
non-conductive pollution occurs. Occasional temporary conductivity
caused by condensation is expected.
IP Rating
IP20 (as defined in IEC 60529).
Cooling Requirements
This instrument relies on the forced air cooling with internal fans and
ventilation openings. Care must be taken to avoid restricting the airflow
around the apertures (fan holes) at each side of the scope. To ensure
adequate ventilation it is required to leave a 15 cm (6 inch) minimum gap
around the sides of the instrument.
CAUTION: Do not block the ventilation holes located on both sides of
the scope.
CAUTION: Do not allow any foreign matter to enter the scope through
the ventilation holes, etc.
Power and Grounding Requirements
The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC
power at 50/60 Hz (+/-5%), or single-phase 100 to 120 Vrms (+/-10%) AC
power at 400 Hz (+/-5%).