FLUKE 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C

User Manual for the Fluke 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C - Page 66

For 7109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

7109A-P-156 photo
Loading ...
Loading ...
Loading ...
6109A/7109A
Operators Manual
62
STATus:PRESet
Set the Operation Status Mask register to 0 and the Questionable Status Mask
register to 0.
STATus:QUEStionable:CONDition?
Return the SCPI Questionable Status Condition register.
(-P models only) The response is 16 if the latest reading of the reference probe
or DUT sensor is out of range.
STATus:QUEStionable[:EVENt]?
Return and clear the SCPI Questionable Status Event register.
(-P models only) The response is 16 if a reading of the reference probe or DUT
sensor was out of range.
STATus:QUEStionable:ENABle <numeric_value>
STATus:QUEStionable:ENABle?
Set bits in the SCPI Questionable Status Mask register.
<numeric_value> is a decimal number ranging from 0 to 255. This determines
which bits of the Questionable Status Event Register affect the QSB summary
message of the Status Byte Register.
SYSTem:BEEP:IMMediate
Make the Product beep.
SYSTem:BEEP:KEY <boolean>
SYSTem:BEEP:KEY?
Enable key beep.
<boolean> is 0 (OFF) or 1 (ON).
SYSTem:COMMunicate:SERial:BAUD <numeric_value>
SYSTem:COMMunicate:SERial:BAUD?
Set the baud rate of the RS-232 port.
<numeric_value> is either 1200, 2400, 4800, 9600, 19200, or 38400.
SYSTem:COMMunicate:SERial:FEED <boolean>
SYSTem:COMMunicate:SERial:FEED?
Turn on serial port monitor.
<boolean> is 0 (OFF) or 1 (ON). When this function is on, the fluid temperature is
transmitted to the serial port automatically at a rate of one reading per second.
The temperature is in °C or °F. The default is OFF.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...