Portable Calibration Baths
Operation
25
Operation
This section explains how to operate the Product after it is set up.
The user interface (UI) and basic key actions are explained in the subsequent
sections.
Display
The display is explained in Figure 13.
1
2
3
4
5
9
6
8
7
icu06.eps
Number Description
Date and Time - The present date and time from the internal, battery-operated clock of
the Product.
Bath Fluid Temperature - The temperature of the bath fluid as measured by the internal
control sensor.
Setpoint - The Product heats or cools to maintain the fluid temperature at the setpoint.
Reference Temperature (-P models only) - If sensors are connected to the input
module and are properly configured, the reference temperature and test sensor reading
show on the home screen.
Test Sensor reading (-P models only) - This is the temperature that the DUT presently
reads.
Control Indicator - Shows Stable when the bath fluid temperature is at the setpoint and
the Product is ready for temperature measurement.
Connection Indicator (-P models only) - The connection indicator shows when a
reference probe is connected and its probe coefficients have successfully been
transferred to the Product.
Recording Indicator - Shows when recording is on.
Heating Status - Shows these states: Off, Heating, Cooling, or Cutout. Cutout indicates
that the bath fluid temperature has exceeded the cutout temperature setting.
Temperature control can only resume after you push and the bath fluid cools
below the cutout temperature.
Figure 13. The Display