FLUKE 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C

User Manual for the Fluke 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C - Page 64

For 7109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

7109A-P-156 photo
Loading ...
Loading ...
Loading ...
6109A/7109A
Operators Manual
60
SOURce:SPOint <numeric_value>
SOURce:SPOint? [MINimum|MAXimum]
Set the temperature setpoint.
<numeric_value> is temperature in °C or °F and depends on the temperature unit
setting. The range depends on the model.
Example:
SOUR:SPO 50.02
Temperature control must be turned on for the setpoint to take effect (see the
command OUTPut:STATe).
SOURce:SPOint:DEFine<index> <numeric_value>
SOURce:SPOint:DEFine<index>? [MINimum|MAXimum]
Set a preset setpoint.
<index> is the preset number. Range is 1 to 8.
<numeric_value> is temperature in °C or °F and depends on the temperature unit
setting. The range depends on the model.
Example:
SOUR:SPO:DEF8 100
SOURce:SPOint:SELect <index>
Set the temperature setpoint to a selected preset.
Temperature control must be turned on for the setpoint to take effect (see the
command OUTPut:STATe).
SOURce:STABility:BEEP <boolean>
SOURce:STABility:BEEP?
Enable the ready beep.
<boolean> is 0 (OFF) or 1 (ON).
SOURce:STABility:CONDition?
Return the ready condition.
The response is 0 (settling or off) or 1 (stable, ready).
SOURce:STABility:LIMit <numeric_value>|DEFault
SOURce:STABility:LIMit?
Set the stability window
<numeric_value> is temperature in °C or °F and depends on the temperature unit
setting. The range depends on the model.
This command is conditionally password protected.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...