FLUKE 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C

User Manual for the Fluke 7109A-P-156 Portable Calibration Bath with Process Electronics, -25°C to 140°C - Page 39

For 7109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

7109A-P-156 photo
Loading ...
Loading ...
Loading ...
Portable Calibration Baths
Menu Guide
35
(Program) New
Create a new program.
See (Program > Select) Edit for a list of program parameters.
Up to eight programs can be defined.
(Program) Delete
Delete the selected program.
(Program) Reports (7109A-P and 6109A-P)
Select a report for view or export.
A list of stored test reports appears. Push or to select a report.
(Program > Reports) View
View the selected report.
(Program > Reports) Export
Export the selected report to a file on a USB memory device.
The memory device must be Linux-compatible and FAT32-formatted.
The filename is created automatically from the program name and the time the
program finished. The extension is “.csv”. The file is placed in a folder on the
memory device that is dependent to the serial number of the Product:
\Fluke\7109_6109\<product serial number>\Report\
(Program > Reports) Delete
Delete the selected report.
Setup
The Setup menu contains operator preferences and instrument settings.
(Setup) Operation
Change settings related to temperature control.
The cutout is a safety device that prevents overheating of the Product
and bath fluid in the event of operator error or product malfunction. If
the fluid temperature exceeds the Cutout Temperature, power is
immediately removed from the heating and cooling devices.
Temperature control can only resume after the operator pushes
and the fluid cools below the cutout temperature. Set the Cutout
Temperature at 5 °C to 15 °C above the maximum temperature at
which the Product is operated and no higher than the maximum safe
temperature of the bath fluid.
Stability Window controls when the control indicator becomes Stable
and the ready indicator turns green. The control temperature must
remain near the setpoint within the Stability Window tolerance for at
least 1 minute.