(Program > Select > Edit) Edit Setpoints
Add, delete, or change program setpoints.
A list of program setpoints is shown. Push or and then to
choose a setpoint to change.
Move to Add and push to add a new setpoint to the end of the list.
(Program > Select > Edit > Edit Setpoints) Insert
Add a setpoint just above the highlighted setpoint.
(Program > Select > Edit > Edit Setpoints) Delete
Delete the highlighted setpoint.
(Program > Select > Edit > Edit Setpoints) Move Up
Exchange the highlighted setpoint with the one just above it.
(Program > Select > Edit) Save
Save the program to memory.
(Program > Select) Options
Change general program and report settings.
If Program Beep is ON, the Product beeps when the program finishes.
In the 7109A-P and 6109A-P, settings to manage reports are also
available.
Report determines whether a test report is created. Readings are
captured at the end of the dwell time and added to the report.
Standard determines the source for the reference temperature. If
Standard is Reference (and a reference probe is connected and
configured) the recorded reference temperature will come from the
reference probe. Otherwise the reference will be the internal control
sensor.
Pass Tolerance specifies the maximum acceptable difference
between the DUT temperature and the reference temperature.