Commands
47
CALCulate:AVERage:CLEAr
Clear the statistical average and standard deviation for fluid temperature,
reference probe, and DUT sensor readings.
CALCulate:AVERage:WINDow <numeric_value>
CALCulate:AVERage:WINDow?
Set the time window in seconds for the moving average and standard deviation.
<numeric_value> is either 10, 60, or 300.
CALCulate1:AVERage[:AVERage]?
(-P models only) Return the moving average of the reference probe readings.
The response is a temperature value in °C or °F and depends on the temperature
unit setting.
CALCulate1:AVERage:SDEViation?
(-P models only) Return the moving standard deviation of the reference probe
readings.
The response is a temperature value in °C or °F and depends on the temperature
unit setting.
CALCulate1:CONVersion:CATalog?
(-P models only) Return the list of supported PRT types for the reference probe.
The response is CVD,ITS90,IEC,RES,NONE.
CALCulate1:CONVersion:DATE <year>,<month>,<day>
CALCulate1:CONVersion:DATE?
(-P models only) Store the given date as the reference probe calibration date.
This command is password protected.
CALCulate1:CONVersion:TYPE?
(-P models only) Set the reference probe type.
<type> is either CVD, ITS-90, IEC, RES, or NONE.
This command is conditionally password protected.
CALCulate1:CONVersion:TYPE <type>
CALCulate1:CONVersion:PARameter:CATalog?
(-P models only) Return the list of characterization coefficients associated with
the selected PRT type.