6109A/7109A
Operators Manual
40
Bath Calibration is the date the Product was calibrated.
TEMP1 adjusts the temperature accuracy at the low end of the Product’s
temperature range.
TEMP2 adjusts the temperature accuracy at the middle of the Product’s
temperature range.
TEMP3 adjusts the temperature accuracy at the high end of the Product’s
temperature range.
Cutout Mode temporarily disables the software cutout to facilitate testing
the hardware cutout. The setting is always Normal after power on.
Typical Stir Power is a reference used to determine whether the load on
the stir motor has changed because of increased fluid viscosity (see
Diagnostics). It is up to the operator to set this value to the observed Stir
Power under normal conditions.
(Setup > Instrument > More > Service) Input Calibration (7109A-P and
6109A-P only)
Adjust the measurement accuracy of the Input Module.
This menu requires the password to change any of the settings.
Input Calibration is the date the Input Module was calibrated.
REF1C0 adjusts the accuracy resistance measurements at 0 .
REF1C100 adjusts the accuracy of resistance measurements at 100 .
TCC0 adjusts the accuracy of the thermocouple measurements at 0 mV.
TCC100 adjusts the accuracy of the thermocouple measurements at
100 mV.
TCCRJ adjusts the accuracy of the thermocouple reference junction.
mAC4 adjusts the accuracy of current measurements at 4 mA.
mAC22 adjusts the accuracy of current measurements at 22 mA.
(Setup) Probe (Models 7109A-P and 6109A-P)
Configure temperature probes connected to the Input Module.
The Probe screen shows the present configuration.
(Setup > Probe) Reference Probe
View or change the reference probe configuration.
If the reference probe connector contains a programmed memory device
(INFO-CON), the probe coefficients are automatically transferred to the
Product when the reference probe is plugged in. If the probe does not
contain programmed memory, the probe coefficients must be entered
manually.