6109A/7109A
Operators Manual
14
Other bath fluids can be used, but the temperature range may be more limited.
Temperature stability and uniformity also varies when other fluids are used. As a
rule, bath fluid viscosity should be no greater than 50 centistokes at any
operating temperature. See Fluid Selection for more information.
Silicone fluid expands as it is heated. This can cause the bath fluid to spill out of
the tank. Use the optional overflow kit or fill the tank to a lower level. The stir
guard has marks MAX and MIN to show the proper fill levels. Fill the tank to the
MIN level if the plug is left in the overflow tube. Fill to the MAX level if the
overflow kit is used. Table 3 gives the recommended fill levels as measured from
the bottom of the tank.
Table 3. Fluid Fill Levels
Overflow Kit Installed Overflow Tube Plugged
MAX level, 154 mm MIN level, 128 mm
Note
Remove the plug from the overflow tube if the tank is filled to
maximum depth.
To add bath fluid:
1. Remove the tank cover.
2. Pour the bath fluid in through the tank opening.
3. Replace the cover and tighten its thumb screws.
4. Clean up any drips or spills. Do not get fluid on the control panel.
Attach the Probe Access Cover
Keep the tank covered when you operate the Product to ensure temperature
stability, uniformity, and accuracy. Use the provided probe access cover. The
cover holes fit many types of temperature sensors. Use the thumb screws to
fasten the cover to the top panel. Make sure the rubber gasket is properly
seated.
Connect to Mains Power
Warning
To prevent possible electrical shock, fire, or personal injury,
connect the factory-supplied three-conductor mains power cord
to a properly-grounded power outlet. Do not use a two-
conductor adapter or extension cord, as it will break the
protective ground connection.
Use the provided mains power cord to connect the Product to a 150 V ac or
230 V ac outlet (this depends on the model). The circuit, outlet, and mains power
cord must all be rated at 115 V ac, 10 A at 230 V ac.