FLUKE 1738/B Advanced Three-Phase Power Logger

User Manual for the Fluke 1738/B Advanced Three-Phase Power Logger - Page 81

For 1738/B.

PDF File Manual, 84 pages, Read Online | Download pdf file

1738/B photo
Loading ...
Loading ...
Loading ...
1736/1738
Users Manual
76
Positioning error with position of conductor in the probe window. (See Table 11.)
Table 11. Flex Probe Window
External magnetic field rejection in reference to external current (with cable >100 mm from the head-coupling
and r-coil) ................................................................40 dB
Phase shift...............................................................<
±0.5°
Bandwidth................................................................10 Hz to 23.5 kHz
Frequency derating .................................................I x f 385 kA Hz
Probe Window iFlex1500-12, iFlex3000-24 iFlex6000-36
A ±(1 % of reading + 0.02 % of range) ±(1.5 % of reading + 0.03 % of range)
B ±(1.5 % of reading + 0.02 % of range) ±(2.0 % of reading + 0.03 % of range)
C ±(2.5 % of reading + 0.02 % of range) ±(4 % of reading + 0.03 % of range)
iFlex1500-12: 88mm
iFlex3000-24: 185mm
iFlex6000-36: 282mm
C
B
A
1/31/31/3
w ww. . com
information@itm.com1.800.561.8187
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com