9100S Dry-Well Calibrator
Calibration Parameters
28
new value. Pressing “EXIT” causes the parameter to be skipped ignoring any changes
that may have been made.
7.12.1
6.12.1 Hard Cutout
This parameter is the temperature above which the unit shuts down automatically.
The value of this parameter is set at the factory to approximately 400°C and cannot be
changed by the user.
The hard cutout parameter is indicated by,
CtOut Flashesforonesecondandthenthehardcutoutsettingis
displayed
400.0 Currenthardcutoutsetting
Press “EXIT” to continue to the next parameter.
6.12.2 R0
This probe parameter refers to the resistance of the control probe at 0°C. The value of
this parameter is set at the factory for best instrument accuracy. The value ranges from
95 to 105. For values greater than 100.000, the display does not show the hundreds
placement. For values less than 100.000, the display shows the entire value. The R0
parameter is indicated by,
r0 FlashesforonesecondandthentheR0settingisdisplayed
00.014 CurrentR0setting(100.014)
To change the R0 setting, press the up or down arrows.
99.999 NewR0setting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewR0setting
6.12.3 ALPHA
This probe parameter refers to the average sensitivity of the probe between 0 and
100°C. The value of this parameter is set at the factory for best instrument accuracy.
aLpha FlashesforonesecondandthentheALPHAsettingis
displayed
38530 CurrentALPHAsetting