FLUKE 9100S-D-156 Dry Block Calibrator with Block D

User Manual for the Fluke 9100S-D-156 Dry Block Calibrator with Block D - Page 14

For 9100S-D-156.

PDF File Manual, 41 pages, Read Online | Download pdf file

9100S-D-156 photo
Loading ...
Loading ...
Loading ...
11
Quick Start
Setting the Temperature
3 Quick Start
3.1 Unpacking
Unpack the dry-well carefully and inspect it for any damage that may have occurred
during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9100S Dry-well
Power Cord
Users Guide with Report of Calibration
RS-232 Cable
9930 Interface-it Software and Users Guide
3.2 Set-up
Place the calibrator on a at surface with at least 6 inches of free space around the
instrument. Always leave enough clearance in front of the instrument to allow for
safe and easy insertion and removal of probes. The prop stand may be swung down to
raise the front of the instrument from a horizontal position. Plug the power cord into a
grounded mains outlet. Observe that the nominal voltage corresponds to that indicated
on the calibrator.
Turn on the power to the calibrator by toggling the power switch on. The fan should
begin quietly blowing air through the instrument and the controller display should
illuminate after 3 seconds. After a brief self-test the controller should begin normal
operation. If the unit fails to operate please check the power connection.
The display should show the well temperature and the well heater will bring the
temperature of the well to the set-point temperature.
After using the calibrator, allow the well to cool by setting the temperature to 25°C
and waiting 1/2 hour before turning the instrument off.
3.3 Power
Plug the instrument power cord into a mains outlet of the proper voltage, frequency,
and current capability. Refer to Section 3.1, Specications, for the power details. Turn
the instrument on using the switch on the rear panel. The instrument will turn on and
begin to heat to the previously programmed temperature set-point. The front panel
LED display will indicate the actual instrument temperature.
3.4 Setting the Temperature
Section explains in detail how to set the temperature set-point on the calibrator using
the front panel keys. The procedure is summarized here.
1. Press “SET” twice to access the set-point value.
2. Press u or d arrow to change the set-point value.