General Operation
Changing Display Units
5 General Operation
5.1 Setting the Temperature
Section explains in detail how to set the temperature set-point on the calibrator using
the front panel keys. The procedure is summarized here.
1. Press “SET” twice to access the set-point value.
2. Press u or d to change the set-point value.
3. Press “SET” to program in the new set-point.
4. Press and hold “EXIT” to return to the temperature display.
When the set-point temperature is changed the controller switches the well heater on
or off to raise or lower the temperature. The displayed well temperature gradually
changes until it reaches the set-point temperature. The well may require 5 to 10
minutes to reach the set-point depending on the span. Another 5 to 10 minutes is
required to stabilize within ±0.1°C of the set-point. Ultimate stability may take 15 to
20 minutes more of stabilization time.
5.2 Changing Display Units
This instrument can display temperature in Celsius or Fahrenheit. The temperature
units are shipped from the factory set to Celsius. There are two ways to change to
Fahrenheit or back to Celsius:
5. Press the “SET” and u simultaneously. The temperature display changes units.
or
6. Press the “SET” key three times from the temperature display to show
Un= C
7. Press the u or d key to change units.
8. Press “SET” to save the setting or “EXIT” to continue without changing the
setting.